10 Cable Fitting
Using Thigh Release Kit (239640) - available separately.
The instructions in this section are for practitioner use only.
1.
Lay cable against socket and locate sleeve inside counterbore in front of chassis (make
allowance for changes in alignment).
2.
Feed inner cable through hole in chassis.
3.
Feed cable through cable clamp and secure, clamping onto tinned area at end of cable.
4.
When fitting cable to thigh release, ensure release lever operates against the patella to
unlock knee as required.
5.
Refer to fitting instructions supplied with Thigh Release Kit.
Knee Locked
1.5
For adjustment of swing control device refer to the PSPC IFU 938057.
Cable Sleeve
Cable
Release
Lever
Cable Clamp
Patella and shin components
not shown for clarity.
13
Knee Unlocked
.
938405PK1/1-0121