5.4 Methanizer Controller Box Connections (Figures 8 and 9)
5.4.1 Be sure the power switch is turned off on the power entry module of methanizer controller box. (O = off,
5.4.2 Connect the H1 output to H1. Align the key on the connector's insert and turn it clockwise ¼ turn.
5.4.3 Connect the TC1 input from the thermocouple to TC1. Push the connector in until it stops.
5.4.4 Plug the VAC cord into the power entry module of methanizer controller box.
5.4.5 Leave the power OFF until you are ready to perform a test.
5.4.6 Connect the power cord to a properly grounded outlet.
5.4.7 DO NOT block the ventilation slots on either side of the methanizer.
Figure 8
H1= Output to heater
TC1= Input from thermocouple
Figure 10: Restek Methanizer Installed in a GC-FID
Figure 9
VAC Input
TC1 Input
H1 Output