3.2 Adjusting Temp A & B High / Low Alarms
1. From the home screen, press SETUP to enter setup menus. If Password Entry
Mode is enable, the controller will prompt for a password. Use the
scroll to the appropriate number and press ENTER to advance the cursor to the
next position.
Note: The default user password is 3 4 5 6.
2. After entering password, the first menu item is Temperature Menus. Press
ENTER.
3. The first menu item is Temp A Menu. For Temp B, press SETUP to scroll to
Temp B Menu. Press ENTER.
4. Press SETUP repeatedly to scroll to the Temp A/B High Alarm setpoint screen
▲/▼
and use the
to select the desire value and press ENTER to save. Press
SETUP to continue.
5. To adjust the Low Temperature Alarm setpoint, use the
desire value and press ENTER to save. Press ESC repeatedly to return to the
home screen.
PN 14910831 Rev A
MVE Variō™ Pro Quick Start Guide
Higher User Level
▲/▼
keys to
Required use
ENTER Password
0000
Press ENTER for
Temperature menus
or press SETUP for
next menu
Press ENTER for
Temp A/B menu
or press SETUP for
next menu
Temp A/B High Alarm
-75.0°C
Use
Press ENTER to save
▲/▼
to select the
Temp A/B Low Alarm
-95.0°C
Use
Press ENTER to save
▲/▼
to
▲/▼
to adjust
▲/▼
to adjust
Page 6