MVE Biological Systems
TM
Alarm Display
ALARM
Lid Open X min
Started at:
01/01/2011 12:00 PM
ALARM
Cooling Time X min
Started at:
01/01/2011 12:00 PM
ALARM
Power Failure
Started at:
01/01/2011 12:00 PM
ALARM
Low Battery
Started at:
01/01/2011 12:00 PM
Communications
Loss Unit 1
Started/Ended at:
01/1/11 12:00
If any alarms occur, contact your authorized MVE Distributors or customer/technical
service.
North & South America
Japan
Asia, Australia, & Pacific Rim
Europe
PN 14910831 Rev A
This alarm is triggered when the Lid Switch Installed setting is set to YES and the freezer lid
remains open for the duration of the time specified by the Lid Open Alarm Delay. Note that
by default, the Lid Switch Installed setting is set to NO.
The Cooling Time Alarm is triggered when the Variō Pro has initiated a Cooling Cycle and the
Chamber Temp fails to reach the minimum chamber temp setpoint in the amount of time
specified by the Cooling Time Alarm Delay.
The Power Failure Alarm is triggered when the Variō Pro is equipped with battery backup
and has been running on battery power for 30 minutes. With typical use, a battery backup
equipped Variō Pro will retain functionality for 48 hours after the loss of its primary power
source.
The Low Battery Alarm is triggered when the voltage of the battery backup powering the
Variō Pro drops below 21VDC.
The Communications Loss screen is displayed in the event that the Display Screen has been
disconnected from the Variō Pro control board. This screen does not necessarily indicate a
loss of system functionality, but it does impair the user's ability to monitor the system
status.