Specifications

This device is powered by an external power supply. When used together, the device
and the power supply are considered ME equipment.
Item
Scales
Accuracy
Resolution (factory default)
Environmental specifications
Protection rating
Weighing range
Readout
Power source
Platform dimensions
Specification
5002, 5122, 5702, 6002
0.4 lb for weights between 4 lb to 40 lb
1% for weights between 40 lb to the weighing capacity of the respective scale
models
0.1 lb / 100 gm
Operation
Temperature : 10 °C to 40 °C (50 °F to 104 °F)
Humidity: 10% RH to 90% RH, non-condensing
Atmospheric pressure: 700 hPa to 1060 hPa
Shipping
Temperature: -20 °C to 50 °C (-4 °F to 122 °F)
Humidity: 15% to 95% RH
Atmospheric pressure: 700 hPa to 1060 hPa
Storage
Temperature -35 °C, humidity uncontrol
Temperature 30 °C, humidity 90%
Temperature 55 °C, humidity 30%
IPX0
Models 5002 and 6002: 4 lb to 880 lb (1.8 kg to 400 kg)
Model 5702: 4 lb to 1000 lb (1.8 kg to 454 kg)
Model 5122: 4 lb to 660 lb (1.8 kg to 300 kg)
Large, bright light-emitting diode display.
Selectable weight displayed in pounds or kilograms.
Indicator light to show selected units.
Battery
6 D alkaline batteries (1.5V x 6)
Power adapter:
Medical grade UL listed external power supply
Nominal input voltage: 100-240V AC ±10%
Nominal input frequency: 50-60 Hz
Nominal input current: 0.4-0.2Arms @ max load
Standby power consumption at U
Nominal output volt age: U
Nominal output current: I
out
5002: 18.25 in. x 19.7 in. x 2.3 in.
: 230V AC : ≤ 0.1W
In
: 12V DC +5% / -5% U
: ≤ 150mVpp
out
Br
: 1500mA
Appendices 25