1. Connect the product and the fitting for wall mounting with moisture-proof
cable H05VV-F.
- Minimum wire cross section 2 x 1 mm², maximum cable length 100 m.
- Alternative: wire cross section 2 x 1.5 mm², maximum cable length
150 m.
2. Strip the wire ends by a maximum of 10 mm.
3. Screw the cable gland with strain relief into the probe head.
4. Connect the wire with the brown insulation or the wire with the black insu-
lation of the cable to one of the two free terminals in the probe head.
5. Connect the wire with the blue insulation of the cable to the remaining free
terminal in the probe head.
6. Tighten the union nut of the cable gland until the cable is sufficiently fas-
tened.
7. Screw the cover with the inserted O ring into the probe head.
8. Connect the wire with the brown or black insulation of the cable to the "+"
terminal in the fitting for wall mounting.
9. Connect the wire with the blue insulation of the cable to the "-" terminal in
the fitting for wall mounting.
10.Use suitable equipment to verify correct electrical installation.
11. Document the installation of the product, see chapter "Certificate of spe-
cialised company".
Mounting
GWG 23-Ro / GWG 23-Wa / GWG 23-T
EN
16