Compex SPORT Mode D'emploi page 23

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 26
GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC IMMUNITY
The Compex stimulator is intended for use in the electromagnetic environment specified below. The customer or the user of
the Compex stimulator should assure that it is used in such an environment.
IMMUNITY TEST
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
a
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the Compex stimulator is used exceeds the applicable RF compliance level
above, the Compex stimulator should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the Compex stimulator.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
b
IEC 60601
COMPLIANCE LEVEL
TEST LEVEL
3 Vrms
150 kHz to 80 MHz
3 V/m
1.4 GHz to 2,7 GHz
10 V/m
26MHz to 1GHz
from structures, objects and people.
ELECTROMAGNETIC ENVIRONMENT -
Portable and mobile RF communications
equipment should be used no closer to any
part of the Compex stimulator, including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2,5 GHz
3Vrms
where P is the maximum output power rating
3V/m
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance
10V/m
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
surveya, should be less than the compliance
level in each frequency rangeb.
Interference may occur in the vicinity of
equipment marked with the following symbol:
EN
GUIDANCE
d = 1.2 √P
in metres (m).
21