It is not advisable to change the alarm setting without the advice of your doctor.

9.5. TOTAL DIFFERENCE - F2

To adjust the temperature difference of MyThermo BW-CX10:
Press the° C /° F button for three seconds until F1 is displayed.
Press the button again for a few seconds to display the F2 menu.
Press the Mode /° C /° F button to change the total deviation from /- 0,1° C / 0.1° F.
Confirm your new SETTING by pressing the° C /° F button for three seconds to display the F3
menu.
In case of doubt, it is advisable to keep the initial parameter at 0,0° C / 0.1° F.

9.6 ACTIVATING OR DEACTIVATING THE AUDIBLE ALARM F3

Press the measurement button to turn on the device.
To display the F1 menu on the screen, press the°C/°F button for 3 seconds. Go to the F3 menu by
pressing the° C /° F button twice for 3 seconds.
Press the°C/°F button to deactivate the audible alarm (0 is displayed on the screen), or on the
MODE button to activate it (1 is displayed on the screen).

9.7 EXIT THE SETUP MENU

Press the° C /° F button for three seconds to confirm changes to the settings and exit the menu.
The LCD screen displays END and powers off.
26
BewellConnect - BW-CX10 - User Manual - 082017

10. TECHNICAL DETAILS AND SPECIFICATIONS

Type of Product
Model
Brand Name
Classification
Temperature sensor
Wavelength
Emissivity
Normal conditions of use
Normal storage conditions
Atmospheric pressure
Battery
Dimensions
Weight
Measurement display
Measurement range:
Repeatability
Measuring Distance
Display Resolution
Accuracy
Bluetooth technology
Electricity consumption
BewellConnect - BW-CX10 - User Manual - 082017
Infrared Medical Thermometer
MyThermo BW-CX10
BewellConnect
Class IIa
IR Sensor
5-14μm
0.95
Ambient temperature
Mode BODY/SURFACE TEMP: 15°C to 40°C /
59°F-104°F
Temperature: -20°C to 50°C / -4°F to 122°F
Relative humidity: 10% ~ 85%
80~105 kPa
DC3.7V (lithium battery)
60,7 x 37,4 x 96,3 mm / 2.36 x 1.46 x 3.78 in
117 g / 0.26 lb
°C/°F
BODY mode: 30° ~ 43°C / 86.0°F ~109.4°F
SURFACE TEMP mode: 0° to 93,3°C / 32°F to 199.9°F
ROOM Mode: 0° C to 50°C / 32° F to 122°F
≤±0.3°C / +/0.3°F
2 - 5 cm / 1-2 in
0,1°C / 0.1°F
For Body mode : See table below
In ROOM mode +/- 1,0°C / +/- 1,8°F
In SURFACE mode ±0,4°C / ±0,7°F
4.0
< 50 mW
27

Publicité