Troubleshooting - Invacare HomeFill II Oxygen Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 135
Indicator Light Color

7.2 Troubleshooting

No indicator lights on when compressor turned On.
RED Light On
1195663-C
Internal failure - contact provider.
Problem
Compressor Status
1. Turn compressor Off (O).
2. Check that power cord is properly plugged into
compressor and wall outlet.
3. Turn the compressor On ( I ).
4. Clean air filters. Refer to Cleaning the Compressor Filter
in the Maintenance section of the manual.
5. If performing Steps 1-4 does not cause any indicator lights
to come on, contact your home health care provider or
Invacare for service.
1. Check the cylinder gauge to see if it is full:
Cylinder full - Remove the cylinder and turn the unit
Off. Perform the Cylinder Prefill Inspection in the
Usage section of the manual with a cylinder that is
not full.
Cylinder not full - Proceed to Step 2.
2. Check that the cylinder is connected properly.
Turn the compressor Off (O).
Remove the cylinder and reinstall.
Turn the compressor On ( I ). If cylinder is not full
and is connected properly, the compressor may have
internal failure. Contact your home health care
provider or Invacare.
Effect
O
flowing to user if concentrator is On.
2
Solution
Troubleshooting
31