Functions; Temperature Display - Grant GD100 Instructions D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

HEATER LAMP
ALARM LAMP
FUNCTION F
POWER

3.1.4.1 Temperature display

The temperature display normally shows the temperature in C of liquid being
controlled. It displays set temperature when the set button "S" is pressed. The
display flashes to indicate that it is in "Set temperature" mode and will return to
actual temperature after a few seconds.
3.1.4.2 Set temperature controls
Press "S" to change display from liquid temperature to set temperature, rotate
the "Navigator" knob to set the required operating temperature. Increasing the
speed of rotation of the knob increases the step size of the set temperature
increment. The display flashes when in the set temperature mode. When the
required temperature is indicated, press "S" again. This value is then stored in
memory and the unit will retain this value after switch off.
3.1.4.3 Set over temperature control (GD120)
The cut-out operating temperature is adjustable. If the liquid temperature rises
above the cut-out set temperature the heater and stirrer are switched off. To
reset the cut-out push the outer ring and release once the bath has cooled
down.
GD100, GD120
Operating Manual
Fig. 1 Front Panel
www.grant.co.uk
CUT-OUT (GD120)
FUNCTION S
17090 Version 15 - June 2009 83S
EN
Page 10