all dimensions in mm (")
metal housing
polycarbonate
housing
INSTALLATION
Mounting of metal and polycarbonate housing
1. Drill the mounting holes according to the mounting template.
For the polycarbonate housing a special mounting set for
installation onto mounting rails is available (refer to "Accessories").
2. The bottom part of the housing is mounted with 4 screws (screw
diameter: < 4.2mm
3. Connection of the transmitter (refer to "electrical connection")
4. Mounting of the middle part and cover with 4 screws (included in
the scope of supply)
Model A (wall mounting)
For mounting template for metal and polycarbonate housing see above.
Transmitter EE23-xAx series are designed for wall mounting.
working range sensor probe:
working range electronics:
The sensor probe must point downwards.
Model B (duct mounting)
For mounting template for metal and polycarbonate housing see above.
Positioning of bore holes top left to center line of the probe:
metal housing:
x = 28.5mm
polycarbonate housing: x = 20.5mm
Transmitter EE23-xAx series are designed for duct mounting.
working range sensor probe:
working range electronics:
The sensor probe must point horizontal or downwards in the duct
channel.
17
; not included in the scope of supply).
(0.2")
-40...60°C
-40...60°C
y = 37.5mm
(1.1")
y = 25.4mm
(0.8")
-40...80°C
-40...60°C
(-40...140°F)
(-40...140°F)
(1.5")
(1")
(-40...176°F)