9. TROUBLE SHOOTING
English
Problem
Power is not ON
Alarm is on
Patient is bottoming out
Mattress form is loose
No air produced from some
air outlets of the air tube
connector
‧ Check if the plug is connected to mains.
‧ Check if the power is suddenly shut down.
‧ Check if the CPR is sealed.
‧ Check if the connection between air tube connector to pump unit is
tightly secured.
‧ Check if all tubing connections along mattress are secured.
‧ Pressure setting might be inadequate for the patient, adjust
comfort range 1 to 2 levels higher and wait for a few minutes for
best comfort.
‧ Check if all the snap buttons or straps of mattress are all securely
fastened.
‧ Check if the mattress is fixed to the bed frame by straps.
‧ This is normal since there is alternating mode. Air outlets take
turns to produce air during their cycle time.
13