Guidance and manufacturer's declaration – electromagnetic emissions
The SIGMA 250 (M2) is intended for use in the electromagnetic environment specified below.
The customer or the user of the SIGMA 250 (M2) should assure that it is used in such environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations/
Flicker Emissions IEC
61000-3-3
Guidance and manufacturer declaration - Electromagnetic immunity
The SIGMA 250 (M2) is intended for use in the electromagnetic environment specified below.
The customer or the user of the SIGMA 250 (M2) should assure that it is used in such environment.
Immunity test
Electrostatic discharge
(ESD) acc. to
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Comment: U
is the a.c. supply voltage prior to application of the test level.
T
44/48
Compliance
Group 1
The SIGMA 250 (M2) uses RF energy only for its internal function. Therefore,
RF-emission is very low and it is unlikely that any interference in nearby
electronic equipment.
Class B
The SIGMA 250 (M2) is suitable for use in all establishments, excluding dome-
stic establishments and those directly connected to the public mains network
that supplies buildings used for domestic purposes.
not
applicable
not
applicable
IEC 60601 test level
± 6 kV contact discharge
± 8 kV air discharge
± 2 kV for mains cables
± 1 kV for input and
output lines
± 1 kV voltage
phase – phase,
± 2 kV voltage
phase – earth
< 5% U
, (>95% dip in U
T
for 1/2 period
40% U
, (60% dip in U
T
for 5 periods
70% U
, (30% dip in U
T
for 25 periods
<5% U
, (>95% dip in U
T
for 5 seconds
3 A/m
MED 113379
Electromagnetic environment – Guidelines
Compliance level
± 6 kV contact discharge
± 8 kV air discharge
not applicable
not applicable
)
not applicable
T
)
T
)
T
)
T
3 A/m
2017-04-13
Electromagnetic environment –
Guideline
Floors should be wood, concrete or
covered with ceramic tiles. If floors
are covered with synthetic material,
the relative humidity should be at
least 30 %.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital environment.