Troubleshooting; Chart Of Medical Device Symbols; Chart Of Electrical Symbols Used - Cuda surgical XLS-300 Manuel De L'opérateur

Illuminateur au xénon
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

11. TROUBLESHOOTING

Problem
The power indicator (refer
to 4.1) is not lit.
The power indicator is lit,
but Xenon lamp will not
ignite.
12. CHART OF MEDICAL DEVICE SYMBOLS USED

13. CHART OF ELECTRICAL SYMBOLS USED

Product labeling
Product label containing model number, serial number, electrical ratings, manufacturer
name, Regulatory marks, and FDA "Rx Only" symbol.
®
LIT037 CUDA
SURGICAL XLS-300
(English)
Solution
A. Check that the AC power cord is properly
connected.
B. Check the fuses. If necessary, replace.
A. Check lamp connection.
B. Check that the lamp door cover is secured.
C. Check hourmeter if lamp hours exceeds the rated lamp life.
D. Replace the lamp (refer to 9.1).
Manufacturer
Date of manufacture (YYYYMM or YYYY)
"Authorized Representative" in the European
Community
Caution, consult accompanying documents
Caution: Federal law restricts this device to sale
by or on the order of a licensed healthcare
practitioner
Consult Instructions for Use
CE mark
AC Current
Equipotentiality
Power on
Power off
Type BF
Rev. N
Date of Revision: 05/14/14
Page 12 of 61