Check the nebulizer and accessories are clean (free of debris), dry and not damaged prior
to use.
Attach the mouthpiece assembly to the handset. Do not use separate mask adapter as
a mouthpiece.
Lift the green medication chamber lid.
Empty the contents of the medication vial into the medication chamber.
• The maximum fill volume is 8ml and this is shown by the word MAX on the protrusion
below the hinge. Do not fill medication above this level.
Close the lid to the medication chamber.
If using a mask, attach the mask adapter to the mouthpiece assembly and attach the mask
to mask adapter.
If you are using the power adapter to power the nebulizer, plug the cable into the socket
on the handset and plug the adapter into the wall outlet.
Press the on/off button on the handset to switch the nebulizer on and begin nebulization.
Check the battery level.
10
• If the LED is SOLID GREEN the battery is charged.
• If the LED is SOLID AMBER, there is enough charge for at least one more treatment.
Please charge your nebulizer after your treatment.
• If the LED FLASHES AMBER and then switches off there is not enough charge to take
a treatment.
• If the LED does not illuminate, please refer to the troubleshooting section of this manual.
Ensure aerosol is coming out of the mouthpiece assembly or mask.
11
a) If using the mouthpiece, hold the handset in your hand and place the mouthpiece
12
between your teeth with your lips sealed around it. Breathe normally through your mouth.
b) If using a mask, hold the handset in your hand and gently press the mask against your
face and breathe normally through your mouth.
During use, some aerosol will be emitted from the back of the mouthpiece assembly.
6
Do not tilt the device in any direction more than 45 degrees during the course
of a treatment as this will prevent the nebulizer from completely nebulizing all the
medication in the chamber.
If you need to take a rest, press the on/off button to stop your treatment. To continue your
13
treatment press the on/off button again.
Your treatment is finished when the nebulizer beeps and the LED flashes. The device will
14
turn off automatically.
Check the medication chamber for residual medication. If there are more than a few drops
15
remaining, press the on/off button again to continue your treatment.
Clean the nebulizer following the cleaning instructions.
16

Charging the battery

Connect the power adapter to the socket on the back
of the handset.
Plug the other end of the power adapter into the
wall outlet.
The LED will PULSE GREEN.
Charge the battery until the LED turns SOLID GREEN
indicating it is fully charged.
Unplug the power adapter from both the handset and
the wall outlet. It is recommended to unplug the power
adapter once the battery is fully charged to preserve
battery life.

Cleaning and maintenance

Cautions
Do not autoclave the mouthpiece assembly or handset.
Do not put any part of the device in a microwave or conventional oven.
Do not immerse the handset in liquid or steam clean.
Never poke or clean the mesh with any sharp objects as this will damage the mesh.
Do not clean the device when in use.
Disconnect device from power supply prior to cleaning.