2.2 ViroSafe® Filter Instructions
Buffalo Filter P/N:
Configuration:
ViroSafe® Filters - Multi Port Filter:
Filter(s):
Particle Size, µm:
Filter Life:
Filter Life Indicator:
ViroSafe® Filter P/N:
Note: Before installing or removing any filter, be sure that the system is turned off.
Filter Installation Instructions:
The installation of the ViroSafe® Filter into the BUFFALO FILTER Turbo Auto-Sense™ Smoke Evacuation
System(s) is quick and simple.
1. Remove the ViroSafe® Filter from the shipping box and discard any protective wrapping. Examine all
filters for damage during shipping and storage. Do not install any filter with visible signs of structural
damage.
2. Insert the ViroSafe® Filter into the filter receptacle. Be sure that the filter is seated completely
against the bottom of the filter chamber.
WARNING: This device is not intended for evacuation of fluid. If fluid is expected to be aspirated
using the ViroSafe® Filter or the BUFFALO FILTER Turbo Auto-Sense™ system, fluid collection
devices must be installed with the vacuum hose assembly. Failure to install a fluid collection device
may cause filter blockage and/or electrical damage.
Filter Removal Instructions:
1. After the ViroSafe® Filter has been exhausted and requires changing, turn the smoke
evacuation system off and disconnect any accessory tubing attached to the filter.
2. Pull the ViroSafe® Filter from the smoke evacuation system and dispose of in accordance with
hospital policy. The ViroSafe® Filter may be disposed of or incinerated.
3. Clean the unit with appropriate germicide prior to re-use. Follow the indicated instructions
for maintenance and installation of a new ViroSafe® Filter.
CAUTION: Using any other filter or accessory not supplied by BUFFALO FILTER may cause
damage to the system and/or cause the system to be inoperable and may void the warranty.
WARNING: The ViroSafe® Filter should be changed when the Filter Life Indicator on the
LCD displays "REPLACE". Failure to change this filter may result in decreased efficiency and
contamination of the electric motor, vacuum pump, and sound absorbing media within the system,
or non-operation of smoke evacuator.
PSWTURBO AS & PSWTURBO AS/220
Portable or tabletop
4-Stage Filtration In One Casing,
(Pre-Filter, ULPA, Carbon, Post-Filter)
ULPA
0.1 to 0.2 Microns At 99.999% Efficiency
Automatic Factory Set Filter Sensor
Timed Replacement
(Call for selection)
Buffalo Filter Turbo Auto-Sense
Operator's Manual:
Page 14