Somnetics Transcend 365 MiniCPAP Guide Rapide page 15

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 27
Transcend 365 miniCPAP Auto
Indicated by a flashing red exclamation point.
During the last therapy session, the heater reached a heat cutoff condition and
turned off for the remainder of the session. The status will be reset when the next
therapy session is started and humidification will be enabled. If the status persists
after subsequent therapy sessions, allowing the system to dry out may correct the
issue. To do this, run the CPAP in Dry Mode. Note: After drying the system,
disconnect and reconnect the power source to ensure the status is reset.
Flashing Red X – No Heater Detected
Indicated by a flashing red X. The humidifier will be disabled. Caused by:
This indicates a corruption, most likely caused by moisture in the CPAP. Allowing
the system to dry out may correct the issue. To do this, run the CPAP in Dry Mode.
Note: After drying the system, disconnect and reconnect the power source to
ensure the status is reset.
If assistance is needed to reset an indicator, contact your medical equipment
provider or Somnetics customer service 877-621-9626.
Combination Indicators
Transcend 365 miniCPAP Auto
The Water and Heater status symbols are not mutually exclusive. It is possible that both
issues may occur and will be shown in combination in the lower right portion of the LCD
screen.
Example:
TM
Quick Guide
TM
Quick Guide
15
ENGLISH
Page 14
Page 15

Publicité

Table des Matières
loading

Produits Connexes pour Somnetics Transcend 365 MiniCPAP

Ce manuel est également adapté pour:

Transcend 365 minicpap auto

Table des Matières