Transcend 365 miniCPAP Auto
Humidifier LCD indicators in Standby
Mode
Water Status Indicators
These indicators will only be displayed if humidifier is set to a level > 0. They will appear
in lower right portion of the Home Screen
Flashing Blue Droplet - Out of Water
Indicated by a flashing blue droplet. Humidifier will be disabled during therapy. Can be
caused by:
No water in the reservoir – only displayed if heater/humidifier is set to level > 0.
If this status appears but there is water in the reservoir, it indicates an error likely
due to moisture in the CPAP. Allowing the system to dry out may correct the issue.
To do this, run the CPAP in Dry Mode. Note: After drying the system, disconnect
and reconnect the power source to ensure the status is reset.