Transcend 365 miniCPAP Auto
Quick Guide
Page 13
Flashing Red Droplet and X – Bad Water Signal Detected
Indicated by a flashing red droplet and red 'X'. The humidifier will be disabled during
therapy.
This indicates a corruption, most likely caused by moisture in the CPAP. Allowing the
system to dry out may correct the issue. To do this, run the CPAP in Dry Mode. Note:
After drying the system, disconnect and reconnect the power source to ensure the status
is reset.
Heater Status Indicators
These appear in lower right portion of the Home Screen (next to the humidification icon)
Flashing Red Exclamation (!) – Heater Over-Temperature
Detected
14
ENGLISH