1-4
STAGE 1 - ATTACHMENT
1. Fully slacken screws A.
Then fit screws B to shank,
after lightly greasing them.
2. Fit the probe to the shank as
shown in the diagram and
visually position the probe
centrally relative to the shank.
Partially tighten screws B
2 - 3 Nm (1.5 - 2.2 lb.ft).
3. Fit the probe/shank assembly
into the machine spindle.
STYLUS ON-CENTRE ADJUSTMENT
Adjusters incorporated in the shank
mounting allow the stylus tip to be
set to correspond with the spindle
centre line.
Stylus tip position is established
using a low force, (less than
0,2N / 0.45lbf) dial test indicator
or setting gauge.
SHANK ATTACHMENT
2,5mm AF
A
B
Shank
Probe
base plate
4mm AF