Immunity test IEC 60601
test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz
to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz
to
2,5 GHz
Note: At 80 MHz and 800 MHz, the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
9.2 Electromagnetic immunity
Compliance
Electromagnetic environment – guidance
level
Not applicable Portable and mobile RF communications
equipment should be used no closer to any part of
the THixyz series, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
3 V/m
Recommended separation distance
d = 1,2
d = 1,2
d = 2,3
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey, a
should be less than the compliance level in each
frequency range b.
Interference may occur in the vicinity of
equipment marked with the following symbol:
- 47 -
80 MHz to 800 MHz
800 MHz to 2,5 GHz