Directions for use
3. Press SEND DATA on the scale to send the current measurement to the computer.

Send measurements to a Welch Allyn device

The scale can connect to a Welch Allyn Connex
Connex
1. Connect the USB Type B male side of the cable to the USB port on the rear of the
scale.
2. Connect the USB Type A male side of the cable into a USB port on the rear of your
device.
3. Press SEND DATA on the scale to send the current measurement to the device.
®
Spot Monitor.
CAUTION Any device connected to the scale must be running on battery
power, a 60601-1-compliant power supply, or a 60601-1-compliant isolation
transformer.
®