10 Scale operation
Audible beeps
When the beeper option is on, the following beeps occur when using the scale.
Note
Sound
One beep
Two beeps
Three beeps
Four long
beeps
Long/short
beeps
Continuous
long beep
Weighing procedure
Make sure to select a consistent time of day to obtain a weight measurement.
Position the scale in a convenient and stable location.
1. Press the power button to turn on the scale.
The scale goes through the startup sequence.
2. Prepare the infant for weighing.
3. Place a scale liner on the scale cradle.
4. Press ZERO to zero the scale.
5. When the scale display reads 0, place the infant on the scale so that the weight is
distributed evenly near the center of the scale.
The beeper option is set to OFF by default. These beeps do not occur
unless beeper option has been enabled in the setup menu. See the
Custom setup section to enable the beeper option.
Meaning
Acknowledgment of a button press.
A zero weight is obtained.
Weight reading is obtained and displayed.
The battery is too weak to operate the scale.
The scale has entered custom setup mode.
There is a problem with the scale. See the troubleshooting section.
There is a problem with the scale. See the troubleshooting section.
CAUTION Replace the cradle after any sign of damage, including small
cracks or crazing. Never place a patient on a cradle exhibiting signs of
damage.
WARNING Patient injury risk. Make sure the scale is properly cleaned
between patients.
WARNING Patient injury risk. Do not reuse the scale liners.
They are single-patient use only.
WARNING Patient injury risk. Never leave the infant
unattended on the scale. Only remove your hands from the
infant for a brief amount of time to allow for an accurate
weight measurement.
Welch Allyn Pediatric/Infant Scale