12 Scale operation
Note
4. After the weight is displayed, press ZERO.
0000 is displayed.
5. Replace the dry diaper with the wet diaper on the center of the cradle.
The scale displays a series of moving dashes until the wet diaper is detected. Once
detected, three beeps sound and the weight displays.
Note
6. The displayed weight is the difference between the wet and dry diaper. Record the
measurement displayed on the scale.
Note
To ensure the most accurate measurements, do the following:
•
Perform the procedure each time a urine output measurement is needed.
•
Do not rely on subtracting a known weight for a dry diaper from the displayed weight
of a wet diaper.
•
Remove any fecal matter from the wet diaper before starting the weighing
procedure. The change in the dry and wet diaper should only reflect urine content.
A series of moving dashes indicates that a tared weight is stored in the scale memory, or
that there is a negative weight on the scale. Press ZERO to clear the tare value and reset
the scale.
Note
Send measurements to a computer
The scale can connect to a computer using the Microsoft
system.
Detailed instructions for this feature are available from Welch Allyn. Go to
www.welchallyn.com/en/other/contact-us.html
1. Connect the USB Type B male side of the cable to the USB port on the rear of the
scale.
2. Connect the USB Type A male side of the cable into a USB port on your computer.
Windows automatically installs the necessary driver.
The beeper option is set to OFF by default. These beeps do
not occur unless beeper option has been enabled in the
setup menu. See the Custom setup section to enable the
beeper option.
The beeper option is set to OFF by default. These beeps do
not occur unless beeper option has been enabled in the
setup menu. See the Custom setup section to enable the
beeper option.
One gram (0.001 kg) is equal to 1 cc of liquid. (1.0 specific
gravity).
When using battery power, the scale does not save tare values after
powering down. See the Custom setup section to change the number of
seconds before the scale turns off when operating on battery power.
CAUTION Any computer connected to the scale must be running on
battery power, a 60601-1-compliant power supply, or a 60601-1-compliant
isolation transformer.