Directions for use
Issue
The reading does not change when
weight is applied to the scale
The scale buttons are not
responding
The error message
displayed
The error message
displayed
The error message
displayed
The error message
displayed
The error message
displayed
The error message
displayed
Possible cause
The load cell transducer, connector, or cable may
be defective
The buttons are damaged
ERR FD
The scale's factory database has failed
is
ERR DB
The scale's application database has failed
is
ERRoR
The scale's flash memory has failed
is
ERR AD
The measurement board cannot be read
is
ERR AV
The measurement board is the wrong version
is
ERR CL
Measurement board error, the weight display is
is
blocked
Storage, cleaning, and maintenance 19
Suggested action
Contact your service department.
Check for visible signs of
damage.
Contact your service department.
Contact your service department.
Contact your service department.
Contact your service department.
Contact your service department.
Contact your service department.
Contact your service department.