4.1 Control unit

The control unit incorporates Newtron® technology patented by SATELEC® S.A. N°US6,765,333B1.
The patented Newtron® technology controls the tips by Cruise Control® , an automatic system for setting the
frequency and power in real time. This ensures that Satelec® tip vibration is gentle, regular and controlled. The
control unit incorporates an ultrasound generator equipped with a piezoelectric command.

4.2 Power configuration dial

To ensure a quality treatment, you will need to use the tips at the power and irrigation flow settings
recommended by Ultradent Products Inc..
The ultrasound power configuration dial ensures:
• Switching On/Switching off of the device (OI symbol).
• Configuration of the operating power: 1 to 20.
The configuration dial has 20 graduations.
• Green: 1 to 9 : very low to low power, used mainly for periodontics.
• Yellow: 6 to 14 : medium power, used mainly for endodontics.
• Blue: 12 to 18 : high power, used mainly for scaling.
• Orange: 17 to 20 : very high power, used mainly for preservation and restoration dentistry.
The ultrasound power configuration dial is not designed to be removed.

4.3 Adjusting ultrasound power

Adjust the ultrasound power using the ultrasound power configuration dial. The ultrasound power must be
adjusted in accordance with the tip used and the required treatment. The operating power of the tips must be
selected in compliance with the Satelec tips color coding system (CCS tips). Details of these indications are given
in the adjustment table available at the address www.satelec.com/documents and also via the Tips recommendations
for use.

4.4 Attach a tip or a file

A tip or a file vibrates correctly when it is perfectly tightened without being forced beyond its stop point.
Tighten it moderately using the wrench provided to ensure optimum ultrasound operation. Over-tightening of the tip
or file can result in breakage of the tip, file or handpiece.
To prevent self-locking of the tip or the file, the latter must be removed after each use.

4.5 Connecting and disconnecting accessories during use

Do not connect/disconnect the cord(s) or the handpiece when the medical device is switched on and
your foot is on the pedal.
Do not tighten or loosen the tips when the handpiece is activated.

4.6 Connecting the handpiece

Check for the absence of signs of moisture at the handpiece connections, and eliminate them if necessary
(wipe and blow using a multipurpose syringe).
Lubricate the irrigation system seal located behind the handpiece with dental instrument lubricant
to extend its effectiveness and prevent leaks.
Connect the handpiece to the sleeve, by avoiding rotation movement.
Install the handpiece on the support.

4.7 Handpiece cord

The cord is only compatible with Ultradent Products Inc. handpieces.
The cord ensures irrigation circulation and electrical connection between the medical device and the
handpiece.

4.8 Light indicator

The light indicator is designed to provide information about the status of the device.
When the light indicator is illuminated, the medical device is on and ready to use.
J58109 • V1 • (14) • 03/2014 • NB79US010B