for prostheses cleaning (or other small items). Each Twin
FingerGuard Basket comes with two (2) basket section
inserts.
Use Directions:
Place Twin FingerGuard Baskets side by side (Fig. 6 & 7),
or use one basket with beaker holder. Twin FingerGuard
Baskets may be used with or without basket section
inserts. Place instruments across ridges in bottom of bas-
ket. Place basket in tank and clean for desired time.
C. Basket Section Inserts:
The UC335 basket section inserts keep instruments sep-
arated by user and procedure, eliminating sorting. This
increases efficiency and reduces the potential for acciden-
tal sticks and cuts since instruments are pre-sorted and
handled less.
Use Directions:
Place locking tabs of adjustable basket section inserts into
basket holes (Fig. 8). Inserts can be placed at any desired
width along the basket. Place as many basket section
inserts as required for your cleaning needs. To remove or
adjust position, simply squeeze center tabs to unlock sec-
tion insert from basket holes.
D. Beaker Holder:
Place the UC330 Beaker Holder on bezel. Place beaker position-
ing rings around the beaker so that at least 2/3 of the beaker is
suspended in the main bath. Clean for desired time.
E. Cassette Racks:
The UC325 instrument cassette basket may be used to hold
up to 3 Large or 6 small standard Hu-Friedy IMS cassets.
Use Directions:
Up to 3 large cassettes may be placed horizontally in each
of the three tiers of the rack. Up to 6 small cassettes can
be placed either horizontally or vertically in the rack (Fig.
9). Placed the filled basket (Fig 10) into the tank containing
either Biosonic General Purpose Cleaner, Biosonic Enzimatic
Cleaner or Biosonic Germicidal Cleaner. Once the cassettes are
cleaned, remove the basket from the tank and tilt back over
the tank to drain. After cleaning rinse the basket and instru-
ments thouroughly under running water. After rinsing the
basket may be placed on the draining tabs on the bezel for
proper draining. (See intrument draining section for details.
Section Insert Support Installation Instructions: Place
the section insert supports in your desired location along the
raised ridge in the bottom of the UC310 Large FingerGuard
Instrument Basket or UC311 Twin FingerGuard Basket. The
section insert supports are inserted into the holes along the
top of the raised ridge with the tapered end facing upward
toward the top of the supports. Position the supports at
equal distance from either end of the basket, along the raised
ridge, to allow the basket section insert to rest squarely in the
basket. Slide basket insert into the slots of the section insert
supports.
VI. Environmental Conditions
• Intended for indoor use.
• Maximum altitude 2000 m.
40°C (104°F)
5°C (41°F)
80%@31°C (88°F)
50%@40°C (104°F)
• Mains supply voltage fluctuations not to exceed ± 10%
of the nominal voltage.
• Transient over voltages category II.
• Pollution degree 2.
• Class I equipment.
EMC Class A: Suitable for use in all establishments other than
domestic and those directly connected to a low voltage power
supply network which supplies buildings used for domestic
purposes.
VII.
Cautions and Warning
NOTE: Failure to comply with these Caution and Warning statements
may result in injury to personnel or equipment damage.
This unit must be properly grounded. Failure to
properly ground unit may result in serious injury.
Do not place hand or fingers into ultrasonic tank
while unit is in operation.
Do not use flammable liquids with BioSonic units.
WARNING!
Keep unit covered to reduce airborne contami-
(Failure to comply
may result in per-
nants.
sonnel injury and/
or equipment dam-
Unspecified or improper use of this product may
age.)
impair safety protection.
Do not operate the unit without sufficient solution
in the tank.
Do not use water alone as the carrier bath. Water
is not a satisfactory coupling agent unless a wetting
agent is used. For best results use BioSonic General
Purpose Cleaner, BioSonic Enzymatic Cleaner or
BioSonic Germicidal Cleaner solutions.
Do not place objects directly on bottom of the tank.
They may block cavitation and cause buildup of heat
in the transducer resulting in premature failure. Use
only BioSonic accessory baskets, cassette racks or
CAUTION!
beaker to hold items to be cleaned.
Ultrasonic cleaning of gemstones, pearls and jewelry
is not recommended as they may be damaged by
ultrasonics.
Do not restrict air flow. Provide adequate ventilation.
Do not subject unit to shock or impact.
Do not immerse the unit in water.