Calibration
1.
Remove the Philips screw on the front of the TX-12-HS. Pull the front cover of the unit off.
2.
Assemble the 25 ppm gas cylinder and regulator together.
3.
Check the pressure gauge on the regulator. If you have 25-psi or less you will need to replace the gas canister.
4.
Place the test Hood from the regulator over the gas sensor.
5.
Push Next 3 times to get to the CAL menu then push Enter. Press Next once to get to Span mode then press Enter,
the display will flash back and forth between gAS and the current gas level.
6.
Start applying 25 ppm gas to the gas sensor by turning on the valve on the regulator.
Note: The sensor will look for the gas for 90 seconds. If no gas is applied or detected in that time, the display will
return to CAL.
7.
When the sensor detects the gas, the display will flash back and forth between the gas concentration and SPn, then
the calibration will progress and the display will show the gas level for a maximum of 165 seconds.
8.
When the calibration is successful, the display will flash back and forth between the gas concentration and PAS, then
the display will show the calibration gas level and the calibration is done.
9.
If the calibration fails, the display will flash back and forth between the gas concentration and FAil (fail). If this occurs,
check the pressure gauge on the regulator. If the pressure is less than 25 psi the flow of gas may not be adequate to
properly calibrate the unit. If there is proper pressure in the cylinder repeat steps 4 through 6. If the unit fails to
calibrate twice contact Technical Assistance: 1-877-367-7891
10.
Once the calibration has passed, remove gas and disassemble the cylinder and regulator.
11.
Re-assemble the TX-12-HS (make sure the LED is aligned with the front case hole). Calibration is complete
12.
See Calibration Flowchart on the inside of the housing.
23