12. MESSAGER OF THE ALLARM AND ERROR DISPLAYS
FAILURE
1. The 2 LEDS
CYCLE ON
END OF CYCLE
blink at the same time.
2. The 4 LEDS
CYCLE ON
END OF CYCLE
121°C
134°C
blink at the same time.
3. Pressure increases very slowly.
MEDILINE ITALIA S.r.l. DO REFUSE ALL RESPONSIBILITY FOR TECHNICAL INTERVENTIONS
CARRY OUT BY NONAUTHORIZED PERSONNEL
Current missing during the sterilisation cycle.
Repeat the operations described on par. 10 and repeat the
sterilisation cycle.
The autoclave has not reached the sufficient pressure.
Before starting the cycle, make sure you have inserted water.
1. The heating element does not heat; in this case, call the
technical service.
2. The safety valve shows leaks.
4. Outlet valve releasing too much steam.
1. The outlet valve shows leaks.
2. Water outlet solenoid valve defect.
3. Supporting level too tilted forwards so insufficient water enter in
the chamber.
4. Leak of the door seal.
CAUSE