Technical Specifications
Scale Indicator
Input signal range:
Sensitivity:
Internal Resolution:
Display Resolution:
System Linearity:
Loadcell excitation Voltage:
Loadcell
Sensitivity:
Input Resistor:
Output Resistor:
Temperature
Operation:
Storage:
Humidity:
Power
Battery:
When the voltage of battery is below 5.7V, the "Low Bat" annunciator will be lighted, plug in AC adapter to charge
the battery. When "Lo.bAt" and actual weight is displayed alternately, this indicates the voltage of battery is below
5.5V and the scale will be turned off in two minutes automatically.
AC Adapter:
10.5-12V
600mA, with central positive, 5.7 VDC -9 VDC
DC
Error Messages
Error Message
0¯ ¯ ¯ ¯:
0_____:
_____:
¯ ¯ ¯ ¯:
CAL-Er:
Lo.bAt:
EEP.E0
EEP.E1
EEP.E2
0mV ~ +30mV
>0.2uV/grad
Approximately 520,000 counts
Can be selected between 500-100,000
Within 0.01% of FS
+5 V
(MAX current: 85mA)
DC
0.3mV/V --- 3mV/V (must be fit to >0.2uV/display grad.)
≥60 Ω
<10 KΩ
o
o
5
C - 35
C
o
o
10
C - 70
C
≤95%RH (no condensation)
6V4AH lead acid battery, 30 hrs continuous use
Definition
Weight above range for calibrated
zero point.
Weight below range for calibrated
zero point.
Indicates an under range condition
Capacity exceeded
Calibration error
Low Battery
EEPROM can't be accessed
Configuration settings have
changed and not been stored
P7-P9 settings exceed scale's
normal range
Required
Solution/Troubleshooting
Remove load before zeroing
Or
Recalibrate the scale.
Remove load before zeroing
Or
Recalibrate the scale.
Remove all loads and zero the
scale.
Remove the load and try again. If
the load is great than 6 kg a scale
with a larger capacity is required.
Restart calibration
Recharge the battery. Upon initial
use, it is recommended to charge
battery for 8 hours prior to use.
Replace S100 Indicator
Calibrate the scale to store
settings
Reconfigure P7-P9 setting