Service Configuration Settings
The scale is configured from the factory to match the specified settings for each unit, as defined by the product
specifications and sales brochure. Modification of the setting can be accomplished by altering user configuration
settings P11-P19. Caution: Configuration of the settings of your scale should be accomplished by a trained
service technician to ensure proper operation and accuracy. Configuration is not covered under warranty.
Setup Controls
Hold
Print
kg
Moves flashing digit
Config
Avail.
Menu
Settings
P11
0,1,2
P12
0-7
P13
00 - 15
P14
0, 1, 2
P15
0, 1, 2
P16
0-8
P17
0-3
P18
0 - 9
P19
0-9
Units
Change flashed digits
Default
Definition
2
Units key
configuration
7
Power On zero
range (full
scale)
3
Zero button
range (Full
scale)
0
Signal within
power on zero
point range
1
Signal not
within power on
zero point
6
Zero tracking
2
Data Filter
1
Weight stability
1
Overload range
Full scale
Tare
kg
Saves data and move down
to next Parameter setting
.
Detailed Setting
0 = KG only ; 1 = LB only ; 2 = units key active KG
and LB
0 = +/- 1% ; 1 = +/- 2% ; 2 = +/- 5% ; 3 = +/- 10% ; 4 =
+/- 20% ; 5 = +/- 50% ; 6 = +/- 100% ; 7 = no limitation
(00) = +/- 1%
(06) = +/- 20%
(01 )= +/- 2%
(07) = +/-no limit
(02 )= +/- 3%
(08 )= + 1%
(03 )= +/- 4%
(09) = + 2%
(04) =+/- 5%
(10 )= + 3%
(05 )= +/- 10%
(11) = + 4%
0 = current weight ; 1 = calibration zero ; 2 = power off
zero point
0 = current weight ; 1 = calibration zero ; 2 = power off
zero point ; 3 = continuously display "0
0 = 0d AZT off ; 1 = +/- 0.25d, 2 = +/- 0.5d ; 3 = +/- 1d;
4= +/- 1.5d ;5= +/- 2d ;6 = +/- 3d; 7= +/- 4d; 8 = +/- 5d
0 = very weak; 1 = weak; 2 = standard; 3 = strong
0 = +/- 0.5d ; 1= +/- 1d; 2= +/- 1.5d ; 3= +/- 2d; 4= +/-
3d ;5= +/- 4d ;6 = +/- 5d; 7= +/- 6d; 8 = +/- 7d; 9 = +/-
8d
0 = 0% ; 1 = +9d ; 2 = 101% ; 3 = 102% ; 4 = 405% ;
5 = 110% ; 6 = 120% ; 7 = 150%; 8 = 200%; 9 = no
limitation
On/Off
Zero
Exits setup mode
(12) = + 5%
(13) = + 10%
(14) = + 20%
(25) = + no limit
- - - - -
"