Connector P8
The numbers listed below refer to the 13 available P8 connections (from left to right) in
1. RS-485 (signal A)
2. RS-485 (signal B)
3. PROFIBUS-DP (GND)
4. PROFIBUS-DP (+ 5 V)
5. PROFIBUS-DP (signal -)
6. PROFIBUS-DP (signal +)
7. PROFIBUS-DP (signal RTS)
Connector P3
Network and access point security is the responsibility of the customer that uses the wireless
instrument. The manufacturer will not be liable for any damages, inclusive however not limited to
indirect, special, consequential or incidental damages, that have been caused by a gap in, or
breach of network security.
Ethernet RJ 45. Connect the instrument to the local network by passing an ethernet cable through
the ethernet cable gland (gland location illustrated in
P3 connector illustrated in
4.7.4 Measurement board
The different measurement boards for the EC and TC sensors are illustrated in
Figure
7. The type of board is easily identified by the color of the J8 connector. For EC boards this
connector is colored orange, and for TC boards it is colored black.
It is extremely important that sensors are connected to the correct measurement board. Connecting
a TC sensor to an EC measurement board (and vice versa) will cause irreparable damage to the
measurement board.
N O T I C E
Figure
4.
N O T I C E
Figure 5 Connector P8
8. Not used
9. Not used
10. Not used
11. System alarm relay (N.O.)
12. System alarm relay (N.C.)
13. System alarm relay (Common)