Functions Indication Table; Function Description - IDEAL 61-737 Manuel D'utilisation Et De Sécurité

Pince ampèremétrique 400 a ca
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Functions Indication Table

Function
Auto Range
LCD
LCD Backlight
High Voltage
Alarm
Regular Prompt
Over Range
Indication
Description
Auto detects and displays most pertinent range for measured value.
Total of two LCDs (front and bottom). Bottom LCD displays same numbers
as front display for all functions, displays a "-" symbol for all negative
readings, displays "AC" for alternating current or "DC" for direct current
and displays "Amps" for current measurement.
White backlight for both front and bottom display.
1) Effective setting: ACV / DCV.
2) In ACV/DCV, voltage measurement value: ACV/DCV ≥30V, LCD
displays high voltage alarm lightning symbol
"
" Simultaneously the NCV LED is on and beeping lasts
for 1 second and then goes silent during measurement.
3) In voltage measure, the voltage data is out of range and
LCD displays "OL", and high voltage alarm lightning symbol
"
" Simultaneously the NCV LED is on and beeping lasts
for 1 second and then goes silent during measurement.
1) When turning the dial switch to any setting position except OFF, the
buzzer will beep one time and NCV LED flashes one time.
2) When the button selection is valid, the buzzer will beep one time; When
the button is invalid, the buzzer will beep twice
3) About 1 minute before the automatic shutdown, the buzzer will beep 5
times continuously, and 1 long beep before the unit shuts down.
4) When the automatic shutdown function is cancelled, the buzzer will
beep 5 times when it reaches the APO time setting.
LCD displays "OL" when over range is encountered.
20

Publicité

Table des Matières
loading

Table des Matières