CARING FOR YOUR TRANSCEND AUTO AND COMPONENTS
Drying Mode
Drying mode should be used daily to dry the inside of your Transcend after use.
Caution:
Do not attempt to use Transcend while in Drying mode. The device will not provide
prescribed therapy pressure while in Drying mode.
1.
Set the device on a dry, stable surface. A cloth surface is not recommended as it may get soiled.
2.
Make sure the device is in Standby mode.
3.
To enable Drying mode, press and hold the Ramp button and then press the Power button.
4.
Low pressure airflow will begin and last for 30 minutes until the drying sequence is completed. The
device will then return to Standby mode. To end drying mode before the 30 minute sequence is
completed, press the Power button. The device will return to Standby mode.
Caution:
Drying mode will dry the inside of your Transcend. Do not attempt to clean the inside of the
Transcend device.
Cleaning the Exterior
Clean the exterior of your Transcend device as needed. To clean, wipe the surface with a lint-free cloth or
paper towel that has been dampened in water containing mild dishwashing detergent*. Following
cleaning, wipe again with a cloth or paper towel moistened with clear water. Allow to dry.
Caution:
Never submerge Transcend in water. Do not allow soap or water to enter the device.
Cleaning the Transcend Components
The interface (mask) should be washed weekly. The air supply tube (hose) should be
washed daily.
1.
Completely disconnect the components.
2.
Hand wash in warm water with a mild dishwashing detergent*.
3.
Rinse thoroughly.
4.
Allow to air dry. Components should be completely dry before reconnecting.
5.
Reconnect the components and reattach to Transcend.
Caution:
Do not put any Transcend components in the dishwasher or laundry machine.
* A solution of 5% mild liquid detergent in distilled water (1.6 fl oz liquid detergent per quart of distilled
water) is recommended.
TRANSCEND AUTO QUICK GUIDE
•
ENGLISH
•
PAGE 9