Medtronic MiniLink REAL-Time Mode D'emploi page 9

Masquer les pouces Voir aussi pour MiniLink REAL-Time:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 19
Disconnecting from sensor
1.
Hold the MiniLink as shown, and pinch the
flexible side arms of the sensor between your
thumb and forefinger.
2.
Gently pull the MiniLink away from the sensor
assembly.
MiniLink tester
The blue tester is used to test the MiniLink and to make sure it is
working.
Connecting the tester
1.
Hold the MiniLink and the tester as
shown. Line up the flat side of the tester
with the flat side of the MiniLink.
2.
Push the tester into the MiniLink until the flexible side arms of the
tester click into the notches on both sides of the MiniLink.
3.
Within 20 seconds the green light on the MiniLink will flash for about
10 seconds when properly connected.
4.
Check the sensor icon on the monitor to ensure that the MiniLink is
sending a signal (see the monitor user guide).
Disconnecting the tester
1.
Hold the MiniLink body as shown and
pinch the side arms of the tester.
2.
With the tester arms pinched, gently pull
the MiniLink away from the tester.
- 7 -

Publicité

Table des Matières
loading

Table des Matières