Air Express
13.0 Troubleshooting
Symptom:
Mattress or overlay does not inflate or
is too soft.
Loss of Pressure Alarm is sounding.
Audible Power Fail Alarm is sounding.
14.0 Specifications, Control Unit
Specification
Enclosure Dimensions
Weight
Power Cord
Overcurrent Protection
Input
Operating Ambient
Temperature Range
Classification
Electromagnetic
Compatibility
Corrective Action
Make certain the air hoses are not kinked, cracked, or split. Verify
the hose couplings are fully engaged to Control Unit. Check that
the air cells are not damaged. Verify that the Control Unit is ON
and that air is coming from either connector outlet.
Additionally on the mattress, verify that all 20 air cells are
connected to the air manifold within the mattress.
Power has been disconnected from Control Unit while in an
operational mode. To restore power, reconnect the cord.
Detachable 4.6m (15'), 18 AWG minimum, with ground wire
Class II grounded equipment not suitable for use in the presence of a
flammable anesthetic mixture with air or with oxygen or nitrous oxide.
IPXO, enclosed equipment without protection against ingress of water.
(CISPR 11 Classified as Class B,
10
C1000DF
31cmW x 23cmH x 15cmD
(12"W x 9"H x 6"D)
2.4kg (5.3 lb)
Primary
1.0A, 250V, T
230 VAC, 50Hz
<37W
16 to 32°C (60 to 90°F )
Type BF equipment
Double insulated
IEC, EN 60601-1
Continuous operation
Meets EN60601-1-2:2001.
Group 1 ISM equipment)