8.3.4.2
Configuration du mode Mesure M3
3 = Limits Mode 3 sert à configurer le module de mesure M3.
Exemple pour CS 223X :
Contamination Code Selection
0=SAE , 1=ISO
current setting: 0
new setting (press Return only to keep setting): 1 'Return'
current setting : 1
lower limit 1:
current setting: 5
new setting (press Return only to keep setting): 16 'Return'
current setting : 16
lower limit 2:
current setting: 5
new setting (press Return only to keep setting): 14 'Return'
current setting : 14
lower limit 3:
current setting: 5
new setting (press Return only to keep setting): 12 'Return'
current setting : 12
(Dans l'exemple, l'objectif de pureté pour le mode M3 a été modifié de SAE 5A, 5B, 5C à
ISO 16/14/12)
8.3.4.3
Configuration du mode mesure M4
4 = Limits Mode 4 sert à configurer le mode de mesure M4.
Contamination Code Selection
0=SAE , 1=ISO
current setting: 0
new setting (press Return only to keep setting): 1 'Return'
current setting : 1
lower limit 1:
current setting: 5
new setting (press Return only to keep setting): 16 'Return'
current setting : 16
lower limit 2:
current setting: 5
new setting (press Return only to keep setting): 14 'Return'
current setting : 14
HYDAC Filtertechnik GmbH
BeWa CS2000 3117721d fr.doc
ContaminationSensor – CS 2000 Serie
fr
Page 37
22/01/2008