Lifting Height; Accessories; Troubleshooting - human care Altair Manuel De L'utilisateur

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Lifting height

The lifting height corresponds to the difference
between the highest and lowest position the
lift can reach. This corresponds to the length of
the lift belt. The length of the lift belt is 1680
mm/66.1"(300 kg), 2050 mm (150/220 kg).
If you want to extend the lift belt there are
distance straps in various lengths available

Accessories

If the maximum weight for any of the
accessories, e.g. slings, lift bars and any
other accessories is NOT the same as for the
lift, then the lowest max. weight given for
any one product is the one that shall apply.
Always check the label on the lift and lifting

Troubleshooting

Description
Nothing happens when you press the Up
button or the Down button on the hand
control or the control panel
Nothing happens when you press the Up
button or the Down button on the hand
control
Nothing happens when you press the Up
button or the Down button on the hand
control or the control panel.
Nothing happens when you press the Up
button or the Down button on the hand
control or the control panel.
Nothing happens when you press the Up
button on the hand control or the control
panel.
Nothing happens when you press the
Down button on the hand control or the
control panel.
The belt is not fed out when the lift is on
a surface.
The indicator switches from flashing
green or yellow to flashing red.
29
as extras. The lifting height of the lift is not
affected by the length of the distance strap.
The total height of the system consists of the rail
in which the lift is mounted, the rail roller in the
rail, a distance strap if relevant and the lift.
accessories or contact Human Care if you have
any questions or if anything is unclear.
Human Care recommends using only Human
Care slings with Human Care lifts. See our
website for all accessories.
Indicator
Action
Not lit
1) Check that the mains switch is on.
2) The batteries are fully discharged. Charge the lift
immediately
3) The emergency stop is pushed in. Deactivate by turning
the red button clockwise.
Flashing green
Check the hand control. If the hand control is defective it
Flashing yellow
needs to be replaced. Contact a service technician.
Flashing red
Constant red
1) Lift motor is overheated. Wait a while and try again.
2) Lift is overloaded. Lower using the emergency lowering
button
Constant yellow
Charging in progress. Interrupt charging.
Flashing green
The lift is in its maximum upper position. Press the Down
Flashing yellow
button on the hand control or the control panel to lower
Flashing red
the lift.
Flashing green
The lift is in its maximum lower position. Press the Up
Flashing yellow
button on the hand control or the control panel to raise
Flashing red
the lift.
Flashing green
The lift belt has to be pulled upwards at the same time as
Constant green
the Down button on the hand control or the control panel
Flashing yellow
is held pressed in.
Flashing red
Flashing red
The batteries need charging but you can still lift the
patient a few times. The lift should be set to charge as
soon as possible.
ENG
ENG
ENG
www.humancaregroup.com

Publicité

Table des Matières
loading

Table des Matières