5.4 Live image

5.5 Synchronization

5.6 Replacing the uEye USB camera

5.7 Adjusting the new camera

Besides the background image, there is the possibility of
showing a live image (for more information please consult the
operating manual included with your BD SCAN PC ALL IN
ONE).
The live image mainly serves to check if the camera is
functioning and connected with the PC (troubleshooting).
Synchronization can only take place with an active connection
to the Internet (for more information please consult the
operating manual included with your BD SCAN PC ALL IN
ONE).
Should the camera (no. 9) of your device no longer be
functioning properly, it can be replaced.
For this, open the lid (no. 2).
Carefully remove the glass plate (no. 1) from the device.
Screw the defective camera out of its mounting.
The defective camera can be replaced by a new uEye USB
camera.
Screw the new camera into the mounting intended for it.
Open the uEye Demo program on your BD SCAN PC ALL IN
ONE.
In the subfolder uEye, open the folder named "Open and Live
Image".
To adjust the focus of the image displayed, turn the lower
screw on the uEye camera (no. 10).
To adjust the exposure of the image, turn the upper screw of
the uEye camera (no. 10).