Certificat de conformité
Certificate:
70006866
Project:
70109280
4. The MA37 and MA160 include different models designation codes. The model designation code is
MA37-x-y and MA160-x-y. Where x and y represent non safety relevant information as listed below:
"x" is a number (1,2,3,...) and denotes the version (e.g. 1);
-
"y" may be up to five numbers and/or letters or blank and denote special versions like country of
-
destination, special software. The hyphens "-"may be omitted.
APPLICABLE REQUIREMENTS
CAN/CSA-C22.2 No. 61010-1-12
CAN/CSA C22.2 No. 61010-2-010:15
UL 61010-1 (3rd Edition)
UL 61010-2-010 (3rd Edition)
CONDITIONS OF ACCEPTABILITY
The main supply cord set provided with the equipment must be an approved type acceptable to the authorities
(1)
in the country where the equipment is sold. It is the manufacturer's responsibility to provide a detachable
power supply cord compliant with the requirements of sub-clause 6.10.1 and that is suitable for the country of
usage and compliant to local regulations and electrical codes. A power supply cord description is provided in
the List of critical components. The power supply cord shall include a protective earth wire and shall be
positioning so that it cannot touch the hot surfaces of the analyzer.
(2) Intended Use: The moisture analyzer can be used for determination of the moisture content of materials of
liquid, pasty, and solid substances using the thermogravimetric method. The analyzer shall be use only for
this purpose. Any other use outside the limits of the technical specifications is not permitted.
(3) The analyzer must not be operated in potentially explosive atmospheres. Unit is not intended to be used with
flammable liquids or materials. Materials or liquids liable to catch fire or explode shall not be used as samples
for moisture analysis.
(4) Appropriate containers must be used for each type of sample material.
(5) If use with substances that contains poisonous or caustic ingredients, the equipment must be use under a
laboratory hood.
(6) Do not put any flammable substances on, under, or near the moisture analyzer. Minimum clearance spacing
required on final installation:
- 1 m above the analyzer
- 20 cm around the analyzer
(7) When operating the analyzer, suitable protective equipment must be worn in order to minimize dangers to
health.
(8) Moisture analysis applications must be optimized by the user in accordance with local regulations.
If at any time there is a conflict between the system safety provisions and any relevant local (national or
regional) requirements, the local requirements always take precedence.
DQD 507 Rev. 2016-02-18
70
Mode d'emploi de l'analyseur d'humidité MA37
-
Safety Requirements for Electrical Equipment for Measurement,
Control, and Laboratory Use, Part 1: General Requirements
-
Particular requirements for laboratory equipment for heating of
materials
-
Safety Requirements for Electrical Equipment for Measurement,
Control, and Laboratory Use - Part 1: General Requirements
-
Particular requirements for laboratory equipment for heating of
materials
Page 2
Master Contract: 167555
Date Issued: 2017-01-16