•
Press the PUMP CURRENT push-button to enter
the data.
The run up time is factory-set to:
00h
08m
00s.
•
When the last digit is entered, the display shows:
D E L A Y
E V E N
T H R E S H O L D :
where: XXX = YES or NO.
•
Press the CYCLE NUMBER pushbutton and select
YES if relay R2 must operate only after the run-up
time or select NO when the R2 operation is needed
right from start of the turbopump and after the
rotational speed of the turbopump exceeds for the
first time the speed threshold value (see the
preceeding cycle diagram). This function is factory-
set to YES.
D E L A Y
E V E N
T H R E S H O L D :
A F T E R
X X X
A F T E R
Y E S
65
TECHNICAL INFORMATION
•
Press PUMP CURRENT to confirm, and the display
shows:
P U M P
L
I
R E S E T ?
where:
•
PUMP LIFE = is the elapsed operating time range
000 to 99999 hours.
•
RESET XXX = YES or NO.
The controller is factory-set to NO.
•
If YES is selected, the pump life shall be reset to 000.
After selecting YES, press the PUMP CURRENT push-
button to enter the command and the display shows:
R E A D Y
F O R
S O F T
S T A R T
or
P U M P
R E A D Y
S T A R T
and the controller is ready to restart (see paragraph
"Startup").
When PUMP LIFE is reset to 000, the CYCLE number
is also reset to 000.
F E
X X X X X h
X X X
L O C A L
:
P U S H
B U T T O N
NOTE
87-900-907-01 (B)