Stylus Fitting Instructions - Renishaw TS20 Guide D'installation Et D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

1-12
Recover stylus tip
from
broken break stem
Hexagon key
Stylus
Countersunk
tip
Broken
break
Apply to upper flats
stem
To recover stylus tip from a
broken stem, use spanner
on upper flats of stem.
BREAK STEM - Please quote the part no. when ordering a replacement stem
Application
25 mm (tooling)

STYLUS FITTING INSTRUCTIONS

The stylus break stem protects the probe in the event of a collision.
Hexagon key
2,5 Nm (1.85 lbf.ft)
screw
Spanner
Break
stem
Before fitting stylus tip onto new
stem, apply Loctite 242 to thread
of countersunk screw. Use
spanner on upper flats of stem.
Break stem
Application
part no.
M-2008-0333
32 mm (tooling)
Fitting stylus tip
onto
new break stem
torque
Spanner
Apply to upper flats
Break stem
part no.
M-2008-0604
Fitting stylus
onto
probe
Spanner
Stylus tip
Apply to lower flats
with
torque
break stem
1,2 Nm (0.88 lbf.ft)
When fitting stylus to probe,
support mounting with
tommy bar and use spanner
on lower flats of stem.
Break stem
Application
part no.
40 mm (tooling)
M-2008-0605
Tommy bar