CAUTION
Hot surface
Burning hazard
Let the device cool down before maintaining.
CAUTION
Tilting risk
Damage of the device
Secure the device against any sudden translocation during
maintenance.
3 Transport and storage
The device must always be stored and transported in the ori-
ginal transport bag. Operation without transport bag is pro-
hibited. Only transport the device in the designated position
(upright, level). If this cannot be ensured due to logistics (e.g.
transport via shippers or air transport), it's important to com-
pletely empty the condensate trap (versions without condens-
ate trap), as condensate could otherwise flow back into the
gas lines. Purge the gas path with preferably dry ambient air
to allow acidic condensate to escape. Proceed as described in
chapter Condensate.
DANGER
Toxic, corrosive condensate
a) Protect yourself from toxic, corrosive condensate when
performing any type of work.
b) Wear appropriate protective equipment.
c) Please note the national safety rules!
The equipment must be protected from moisture and heat
when not in use. It must be stored in a covered, dry and dust-
free room at a temperature between -20 °C and 40 °C.
Outdoor storage is prohibited. On principle the operator must
meet all applicable standards with respect to preventing dam-
age due to lightning, which could result in sample gas pump
damage.
Storage areas must not contain any equipment generating
ozone, e.g. fluorescent lighting, mercury vapour lamps, high
voltage electrical equipment.
4 Setup and connection
Check the device for damage prior to installation. This/these
could be a damaged housing or add-on components visible
from the outside, such as filter and flow meter. Never use
equipment with obvious damage.
CAUTION
Damage/health hazard due to heater leakage
Check the device for obvious leaks before every use and at
regular intervals. Any leaks must be repaired prior to using
the device. In addition to gas leaks, leaking fluids in particular
can post an electrical and health hazard.
BX464003 ◦ 11/2017

4.1 Installation site requirements

Be sure the equipment is located on a level, solid surface. Also
be sure to comply with the approved ambient temperature.
Do not obstruct the convection of the cooler. There must be
adequate room between the vent and the next obstacle (at
least 10 cm).
CAUTION
Damage to the device
Protect the equipment against dust, falling objects and ex-
ternal impacts.
Stroke of lightning
Outdoor installation is forbidden. As a matter of principle, the
operator must regard all applicable standards according pre-
vention of damage due to lightning, which may otherwise
damage the device.
4.2 Connecting a gas probe
The sample gas probe connects to the DN 6 hose connection
on the device marked IN.
The following devices (e.g. analyser) must be connected to the
DN 4 hose connection marked OUT with a suitable hose.

4.3 Electrical connections

WARNING
Hazardous electrical voltage
The device must be installed by trained staff only.
CAUTION
Wrong mains voltage
Wrong mains voltage may damage the device.
Regard the correct mains voltage as given on the type plate.
The low heat device socket features an on/off switch which
cuts off all poles. This must be set to the zero position prior to
connecting the electrical.
Connect the included low heat device cable to the device and a
suitable voltage source. Please note the correct voltage and
frequency. The type plate contains any deviating specifica-
tions.
PCS.base with condensate trap may be operated on
110-260 V AC, 50/60 Hz voltage. When using the condensate
pump the voltage is 115 V, 60 Hz or 230 V, 50 Hz.
Bühler Technologies GmbH
7