PCS.base

4.4 DIP switch settings

DIP switch
The unit is configured using four DIP switches at the front of
the cooler.
1
ON switch
SW1
0
1
0
1
0 OFF switch
SW2
0
1
SW3
S
Switch, the following numbering of the SWs
0
1
SW4
W
corresponds with the numbering on the DIP
switch.
SW1 / SW2
SW2 SW1 Gas output dew point
0
0
3 °C
0
1
5 °C (factory preset)
1
0
10 °C
1
1
15 °C
SW3 / SW4
SW3 SW4 Delta T control
0
0
Gas output dew point, fixed
0
1
Difference from ambient temperat-
ure approx. 15 °C
1
0
Difference from ambient temperat-
ure approx. 30 °C
1
1
Gas output dew point, fixed
5 Operation and control
NOTICE
The PCS.base is a portable device which must only be oper-
ated inside the original transport bag. The bag must be open
during operation.
Only operate and transport the unit upright. Operating the
device with the cover closed or outside the specifications is
prohibited!
CAUTION
Damage/health hazard due to heater leakage
Check the device for obvious leaks before every use and at
regular intervals. Any leaks must be repaired prior to using
the device. In addition to gas leaks, leaking fluids in particular
can post an electrical and health hazard.
5.1 Switching on the PCS.base
After switching on the supply voltage the cooler starts to cool
the cooling block.
The target temperature is factory preset to 5 °C. The alarm
limit is defined at +5/-2 K.
8
Bühler Technologies GmbH
Green
Red
In-
FF
LED
LED
ternal
status
OFF
OFF
Unit off
ON
OFF
OK
OK (*)
OFF
Flash-
OK
OK (*)
ing
f = 1 Hz
OFF
ON
OK
Error
OFF
Flash-
Error
xxx
ing
f = 5 Hz
OK
No error
Error
Error present
xxx
Status not defined
f =...
LED flashing frequency
(*)
Also applies if no moisture detector connected
If the red LED lights up during operation, please refer to
chapter "Troubleshooting".
5.2 Operating the sample gas pump
DANGER
Toxic, corrosive gases
The measuring gas led through the equipment can be hazard-
ous when breathing or touching it.
a) Check tightness of the measuring system before putting
it into operation.
b) Take care that harmful gases are exhausted to a save
place.
c) Before maintenance turn off the gas supply and make
sure that it cannot be turned on unintentionally.
d) Protect yourself during maintenance against toxic / cor-
rosive gases. Use suitable protective equipment.
The pump will only run after reaching the block temperature.
It is only intended to convey gaseous mediums. It is not suit-
able for liquids.
Operation under primary pressure is prohibited. The flow in-
side the device (operational regulation via flow meter with
built-in needle valve) should always be at least 50 L/h. Throt-
tling greatly will shorten the life of the sample gas pump.
CAUTION
Hot surface
Risk of burns
Housing temperatures may be high during operation.
Allow the unit to cool down before performing maintenance
or repairs.
Tem-
Description
perat-
ure
With the cooler
switched off, the status
output corresponds to
error status.
OK
Normal operation
Error Overload / temperat-
ure outside the target
range
xxx
Moisture penetration
xxx
Various possible
causes, contact Service.
BX464003 ◦ 11/2017