2 Description
The Parker domnick hunter UHPZN2 range of zero nitrogen generators employ robust, field proven technology to produce ultra high purity
nitrogen for GC makeup and carrier gas applications. Flow rates range from 1.0 L/min (0.04 cfm) to 3 L/min (0.11 cfm), with purities <10 ppm.
The generator provides a continuous stream of ultra high purity nitrogen from a single 'plug & play' unit. Models are available with and without an
integral oil free compressor.
2.1 Technical Specification
This specification is valid when the equipment is located, installed, operated, and maintained as specified within this user guide.
Outlet Parameters
Flow
(1)
Set Pressure
(2)
Purity
content
(O
)
2
(3)
Purity (Hydrocarbon Content)
(4)
Dewpoint (@ 5 barg)
Particulates
Inlet Parameters
Integral Compressor
Air Quality
Air Pressure
Minimum Air Temperature
Maximum Air Temperature
Air : N
Ratio (@ 9 barg)
2
Peak Flow (@ 9 barg)
Port Connections
(5)
Air Inlet
(6)
Gas Outlet
Drain
Electrical Data
Supply Voltage
Connection type
Current
(7)
Power
(8)
Fuse
Supply Voltage
Connection type
Current
(7)
Power
(9)
Fuse
(1) If you are using the generator's internal Mass Flow Controller, please note that the pressure at the N2 MFC Outlet port will be approximately 0.5 bar (7.3 psi) below the set
pressure.
(2) The specified purities are referenced to 20
monitored in critical applications.
(3) With a <100 ppm CH
challenge
4
(4) The Pressure Dew Point (PDP) is the temperature at which the water vapour within the pressurised gas condenses in to a liquid.
(5) Using the 1/4Ý BSPP - 1/4Ý tube fittings provided.
(6) Using the 1/8Ý BSPP - 1/8Ý tube fitting provided.
(7) The power consumption in standby mode and economy mode is 3W for all models.
(8) Fuses are Anti surge (T), 250V, 5x20mm HBC, Breaking capacity 1500A @250V, IEC60127, UL R/C.
(9) 8A fuses are Anti surge (T), 250V, 5x20mm HBC, Breaking capacity 1500A @250V, IEC60127, UL R/C. 12.5A fuses are Anti surge (T), 250V, 6.3 x 32mm HBC, Breaking Capacity
1500A @440V, IEC 60127, UL R/C.
UHPZN2
Units
1000
L/min (cfm)
1.0 (0.04)
bar g
(psig / MPa / kPa)
ppm
ppm
o
o
PDP
C (
F)
ISO 8573-1: 2010
-
No
ISO 8573-1: 2010
3.2.1
bar g
(psi g / MPa / kPa)
o
o
C (
F)
o
o
C (
F)
--
24.1
L/min (cfm)
42.0
--
1/4"
--
--
Vac
-
A
2.82
W
632
A
5
Vac
-
IEC 60320 - C14
A
1.3
W
105
A
8
o
C ambient temperature. A reduction in purity may be experienced with increased temperature, it is recommended that the purity is
UHPZN2
1000C
1.0 (0.04)
5
(72.5 / 0.5 / 500)
<5
<0.1
-62 (-79.6)
Class 2
Yes
N/A
9 - 9.9
(130.5 - 143.6 / 0.9 - 0.99 / 900 - 990)
15 (59)
25 (77)
--
--
N/A
1/8"
1/4"
207 - 253 V 50/60Hz
IEC 60320 - C14
4.8
763
8
104 - 127 V 60Hz
IEC 60320 - C19
4.22
473
12.5
UHPZN2
UHPZN2
3000
3000C
3.0 (0.11)
3.0 (0.11)
<10
No
3.2.1
11.1
52
1/4"
1/8"
2.82
632
5
IEC 60320 - C14
IEC 60320 - C19
1.3
105
8
Yes
N/A
--
--
N/A
4.8
763
8
4.22
473
12.5
3