Guidance And Manufacturer's Declaration - Electromagnetic Immunity - Sportec 27512 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 84
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE U T is the a.c.mains voltage prior to application of the test level.
The essential performance of the Sonopuls 190 is as follows: interference-free delivery of ultrasound,
interference-free control of all functions. Uninterrupted operation is not required with the use intended.
Guidance and manufacturer's declaration – electromagnetic immunity
The Sonopuls 190 is intended for use in the electromagnetic environment specified below. The customer
or the user of the Sonopuls 190 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/ m
IEC 61000-4-3
80 MHz to 2,5 GHz
supply lines
 1 kV for input/output
Lines
 1 kV line(s)to line(s)
 2 kV line(s)to earth
<5 %U T
(>95 %dip in U
)
T
for 0,5 cycle
40 %U T
(60 %dip in U T)
for 5 cycles
70 % U T
(30 %dip in U T)
for 25 cycles
<5 % U T
(>95 %dip in U T)
for 5 sec
3 A/m
Compliance
supply lines
not applicable
 1 kV line(s)to line(s)
 2 kV line(s)to earth
< 5% U
T
for 0,5 cycle
< 5% U
T
for 1 cycle
70% U
T
for 25 cycles
< 5% U
T
For 5 sec.
3A / m
Electromagnetic environment –
level
Portable and mobile RF
communications equipment should be
used no closer to any part of the
Sonopuls 190, including cables, than
the recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
3 V
d = [3,5/3]√P
3 V/ m
d = [3,5/3]√P
d = [7/3]√P
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
less than the compliance level in each
frequency range.
occur in the vicinity of equipment
marked with the following symbol:
should be that of a
typical computer room.
Mains power quality
should be that of a
typical computer room.
Mains power quality
should be that of a
typical computer room. If
the user of the
Sonopuls 190 requires
continued operation during
power mains interruptions,
it is recommended
that the Sonopuls 190 be
powered from an
uninterruptible power
supply or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
computer room.
guidance
80 MHz to 800 MHz
800 MHz to 2,5 GHz
a
should be
b