Instructions for use elements e-motion Dental Motor
5 Startup | 5.1 Connector

5.1.3 Connect the motor cord

▶ Plug the motor cord onto the motor cord connector on the rear of the ele-
ments e-motion and screw it tight.

5.1.4 Connect the power supply

Power supply and cords/hoses are on the floor.
Slipping and tripping.
▶ Power supply and cord/hoses need to be set-up/routed appropriately such
that they are not on the floor.
Property damage from non-approved power supply.
Property damage to the product.
▶ Operate the product with the power supply (Mat. no. 1.013.4796) exclus-
ively.
Note
The connection of the power supply must comply with the country-specific
regulations and requirements for medical devices.
Note
The power supply automatically adjusts to the available mains voltage.
Note
The protective earth conductor is used as functional earthing (FE) rather than
as protective earthing (PE).
Note
Use the power supply in dry environments exclusively. Make sure that the
power supply is protected from the ingress of liquids.