Indicated by a flashing red droplet and red 'X'. The humidifier will be disabled during therapy.
This indicates a corruption, most likely caused by moisture in the CPAP. Allowing the
system to dry out may correct the issue. To do this, run the CPAP in Dry Mode.
Note:
After drying the system, disconnect and reconnect the power source to
ensure the status is reset.
Heater Status Indicators
These appear in lower right portion of the Home Screen (next to the humidification icon)
Flashing Red Exclamation (!) – Heater Over-Temperature Detected
Indicated by a flashing red exclamation point.
During the last therapy session, the heater reached a heat cutoff condition and
turned off for the remainder of the session. The status will be reset when the next
therapy session is started, and humidification will be enabled. If the status persists
after subsequent therapy sessions, allowing the system to dry out may correct the
issue. To do this, run the CPAP in Dry Mode.
Note:
After drying the system, disconnect and reconnect the power source to
ensure the status is reset.
Flashing red
exclamation point
Page 12