Transcend 365 miniCPAP Quick Guide
Use the Up or Down buttons to move the humidification level from 0 to 10 (0 = no
humidification, 10 = maximum humidification). You will see the water droplet icons move
to a higher or lower value. Once you reach the desired setting, use the left and
right buttons to toggle to the green check mark, if it's not already highlighted,
then press the Center button to save your settings. To exit out of the humidifier
settings without saving any changes, select the red X by using the left and right buttons
to highlight it, then press the Center button.
Push the Left/back button to return to the Home screen and you will see your new
humidifier setting displayed on the Humidification icon.
Humidifier LCD Indicators in Standby
Mode
Water Status Indicators
These indicators will only be displayed if the humidifier is set to a level > 0. They will
appear in lower right portion of the Home Screen.
Flashing Blue Droplet - Out of Water
Indicated by a flashing blue droplet. Humidifier will be disabled during therapy. Can be
caused by:
No water in the reservoir – only displayed if heater/humidifier is set to level > 0.
If this status appears but there is water in the reservoir, it indicates an error likely
due to moisture in the CPAP. Allowing the system to dry out may correct the issue.
To do this, run the CPAP in Dry Mode.
Note:
After drying the system, disconnect and reconnect the power source to
ensure the status is reset.
Flashing blue droplet
Page 10