Calibration Checks - Testing Instructions; (For Trained Specialist Staff Only) - boso medicus smart CA01 Mode D'emploi

Masquer les pouces Voir aussi pour medicus smart CA01:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Calibration checks - Testing instructions

(for trained specialist staff only)

Calibration checks – every 2 years at the
latest – must be performed by the follo-
wing institutions or persons:
– Manufacturer
– Trade measurements authorities
– Persons who fulfil the requirements of
the legislation governing the operation
of medical devices.
A) Function testing
Function testing of the monitor can
only be carried out on a person or with
a suitable simulator.
B)Testing of Density of Pressure
Circuit
and
Pressure Indication
60
Divergence
from
1. Remove battery.
2. Hold Start button
the same time insert battery.
3. Release Start button
is ready for testing. The displays of SYS
and DIA now show the actual pressu-
re.
4. Prior to testing the density, open the
battery compartment and take the T-
connector off the unit. Wrap the cuff
around a fixed cylinder and connect it
with the T-connector (without deflati-
on valve) to the pump ball and the
unit.
pressed and at
. The unit now

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Medicus smart ca02

Table des Matières